The most common mistake buyers make with the word "tested" is believing it without asking what was measured, on which unit, with which instrument, and what the seller would have done if the number had been bad. A test you cannot inspect is a marketing sentence. We run the units in our own sessions, we do not manufacture or sell hardware, and this page is the protocol we follow: what we check, what we record, what we cannot check, and the rule for a failure.
What we check before a unit is used
Every unit gets an acceptance test before its first session. The test has four blocks. The first is mains safety: earth continuity and insulation resistance for anything that plugs in, plus a visual check of the lead, the plug and the case. The second is output: the driver waveform, the frequency, the duty cycle and the flux density at fixed distances from the emitter. The third is thermal: surface temperature at the point of contact over a full session, with the controller at its normal setting. The fourth is control: the timer, the shut-off and the behaviour when the setting is changed mid-session.
We buy two units of a model when we can, because a single unit tells you about a unit and two tell you something about a model. Each gets a serial number, a commissioning sheet and a reference measurement. The reference is repeated every six months, and the question at that check is not whether the unit still works but whether it has moved: a coil that has drifted ten per cent is still working and no longer the device we tested.
The sheet is deliberately boring. Date, serial, instrument, distance, reading, uncertainty, initials. A log that can be reconstructed by someone else is worth more than a certificate that cannot.
The instruments and what each one reads
Portable appliance tester. It measures earth continuity and insulation resistance at a defined test voltage. Those two numbers decide whether a mains device is safe to plug in, and they have nothing to do with the field. Our pass criteria are the standard ones: earth continuity at or below 0.1 ohm and insulation at or above 1 megohm at 500 volts, with the lead and the connector tested as part of the unit.
Oscilloscope and current probe. The scope shows the drive waveform, the frequency and the duty cycle at the coil. It is how we catch a controller that claims a sine and delivers a square, or a frequency that wanders as the unit warms up. A scope reads the electronics, not the field in the room; the two are related but not interchangeable.
Three-axis gaussmeter on a fixed jig. The jig holds the probe at a fixed distance and angle from the emitter, which is the only way to compare readings over time. We record flux density at the surface, at ten centimetres and at thirty centimetres, on all three axes. The room's background is measured first and subtracted, because a room with wiring is not a zero.
Infrared camera and contact probe. The camera shows the hot spots; the contact probe gives a number at the point where skin would rest. We run a full session at the normal setting and record the maximum surface temperature. A device that is comfortable for ten minutes and hot at forty is a device with a timer problem.
Power meter and sound level meter. Consumption tells us whether the unit is drawing what the label says. Coil whine is not a safety issue, but a unit that buzzes at a frequency you can hear is a unit that will be returned, and it is often a sign of a loose winding.
What we cannot measure
We cannot measure dose in tissue. The field at the surface is not the field at depth, and converting one into the other requires a body model and a calculation that consumer vendors do not publish. At these frequencies the relevant quantity is induced current density, and we do not have it. Neither does anyone selling a mat without dosimetry.
We cannot measure "scalar" as a quantity. There is no unit, no sensor and no calibration procedure for it. What we can measure is voltage, current, flux density, temperature and power, and those are the numbers in the log. A seller who says the output is scalar and therefore cannot be read on instruments is not describing a device we can test.
We cannot measure whether a session works for you. We can measure that the unit is safe, stable and within its reference. We cannot measure sleep quality, pain, tension or the expectation that moves all three. Our tests are engineering quality control. They are not a clinical trial, they do not have a control group, and they cannot separate the device from the ritual.
We also cannot measure the long term. A unit that passes today may drift in two years; a session that feels pleasant may do nothing measurable in twelve weeks. The six-month reference check is our answer to the first problem, imperfect and slow. There is no answer to the second within our reach.
| Test | Instrument | Pass criterion | What it still does not prove |
|---|---|---|---|
| Mains safety | Appliance tester, visual check | Earth at or below 0.1 ohm; insulation at or above 1 megohm | Anything about the field |
| Frequency and waveform | Oscilloscope with current probe | Within one per cent of the stated frequency over a session | Field strength in space |
| Flux density at distance | Three-axis gaussmeter on a fixed jig | Within fifteen per cent of the commissioning reference at 10 cm | Dose in tissue |
| Surface temperature | Infrared camera and contact probe | Below the manufacturer's contact limit after a full session | Long-term exposure |
| Timer and shut-off | Controller log and stopwatch | Off within sixty seconds of the set time | Whether a session suits you |
| Drift at six months | Repeat of the field and frequency tests | Within ten per cent of reference, or retire the unit | Whether the model is effective |
What happens when a test fails
A mains safety failure is the only one that requires no discussion. The unit is unplugged, tagged, removed from the session room the same day and returned to the supplier. It is never used with a customer again, whatever the supplier says and whatever the unit cost. Electrical safety is not a trade-off against convenience.
An output failure is slower. The unit is quarantined, the measurement is repeated with a second instrument, and the probe position is checked before the device is blamed. If the deviation is confirmed, the unit is retired from sessions; if it is within tolerance but has moved, the new reference is recorded and the drift is watched at the next check. A unit that has drifted more than ten per cent from its commissioning reference is not used, because the sessions we describe were priced and designed around the device we tested.
A thermal or timer failure means the controller is replaced if a spare exists, or the unit is retired. We do not defeat a thermal cut-out, extend a timer beyond its rating or run a unit with the case open to keep it cool. A model that fails as a batch is stopped as a model: we do not keep using the good units from a bad batch and hope. If the failure changes what we tell customers, this page is updated with the model and the date, and the people who had sessions on that model are contacted in plain language.
What this protocol does not buy us
It does not make the therapy evidence-based. A clean safety test and a stable field say nothing about inflammation, sleep or pain, and no amount of engineering rigour turns a wellness session into a clinical treatment. It does not replace the manufacturer's declaration of conformity, which is a legal document about the product and not something we can issue on their behalf. And it does not make us independent: we run the units we test, so we have an interest in the answer.
What it buys is a paper trail. If a unit is withdrawn, there is a date and a serial. If a claim about output is challenged, there is a measurement. If we are wrong, the log shows where. That is a lower standard than a trial and a higher one than a testimonial, and it is the standard we can actually meet.
Where scalar energy fits
Buying a device is a decision about money and expectations, not about belief. The sessions are used for relaxation and sleep; that is what you are paying for, and it is worth paying for if it works for you, but it is not a medical device and nobody should sell it as one. The test is simple: fixed cost, fixed number of weeks, a note on how you slept, and a decision at the end. If the seller will not let you run that test, the price is not the problem. The mechanism argument is in scalar energy and inflammation; the substance layer is in best supplements for inflammation.
Testing a device is not the same as proving a treatment. It is the smaller, checkable promise: the box is safe, the output is what we said, the unit that failed is not in the room, and the log says so.